General information
SmartScope comes with a few protocols that essentially dictate which steps are going to be performed during the imaging process and which algorithms are going to be used.
Some protocols come with a -precise
variant. The main differences with their regular counterpart are:
- Eucentric height performed at the medium mag rather than the square mag. This is more precise as the hole pattern from the square mag can lead to erroneous cross-correlation.
- Hole re-centering should be more precise. This allows to start beam-image shift imaging with less initial image-shift.
When setting up a grid, all the protocols are availble via a dropdown menu with a additional default auto
choice. This option follows a set of rule to select the best protocol for the grid and imaging type. Here's a breakdown of these rules:
Note
The auto
will never select the -precise
variant of a protocol. It has to be manually selected.
- SPA-Ptolemy -> Selected for all holey grids if tilt angle is 0 in the imaging parameters.
- SPA -> Selected for holey grid if tilt angle is not 0.
- NegativeStain -> Selected if the grid type is
Negative Stain
orLacey
Advanced - Changing the rules for auto
These rules can be modified in a settings file. More details here